Environmental Testing Catalog Dissolved Cations ASTM D6919
DCLS Method
Dissolved Cations and Hardness Determination by Ion Chromatography
Reference Method
ASTM D6919
Test Description
The test determines cations by suppressed ion ion chromatography. It is also used to calculate the hardness of drinking water.
Sample Collection
The collection is preserved to a pH of ≤2. Cations typically are preserved with nitric acid.
Storage Guidelines
Store at ≤4 degrees celcius, but not frozen.
Testing Group
Inorganics
Analytes
Calcium
Magnesium